Carewell Digital-1106L ECG


Item № ECG-1106L
Six channel digital electrocardiograph

Quick overview

1. 12-lead simultaneous acquisition
2. 6” LCD, rea-time display of 12-lead ECG waveforms
3. Detection and alarm on lead-off and low battery capacity
4. Automatic adjustment of baseline for optimaprinting
5. Built-in ECG simulator for DEMO purpose
6. 3 kinds of operation mode: AUTO, MAN and ANA
7. 3 kinds of filters: HUM, EMG and ADS
8. Power supply: AD, DC, battery
9. 2 types lead mode: Standard and Cabrera
10. 1,000 ECG files can be save to SD card
11. Communication with PC (Option)
 
Carewell Digital-1106L ECG

Details

Functions of ECG-1106L are listed as in below table

MODEL       

AUTO MEAS

AUTO ANA

LCD SIZE

DISPLAY

PAPER WIDTH

INFO

FILE

HELP

RS232

USB

WIRE-

LESS

SD CARD

ECG-1106L

Yes

Yes

6”

12CH

112MM

Yes

Yes

Yes

Yes

Yes

Option

Yes

Standard configuration

  • ECG-1106L main unit
  • Patient cable
  • Chest electrode
  • Limb electrode
  • Thermal paper
  • Paper roller
  • Rechargeable battery
  • Dust cover

Options

  • PCECG-500A Workstation software package
  • Aluminum trolley

SafetyStandards

MDD93/42/EEC

Medical Device Directive

IEC 60601-1: 1988

+A1:1991+A2:1995

Medical electrical equipment - Part 1: General requirements for basic safety and essential performance 

IEC60601-2-25:1993 + A1:1999

Medical electrical equipment - Part 2-25: Particular requirements for the safety of electrocardiographs; Amendment 1 

IEC 60601-1-2: 2007

Medical electrical equipment Part 1: General requirements for safety -2 Collateral standard: Electromagnetic compatibility - Requirements and tests

EN 60601-1-4

Medical electrical equipment - Part 1-4: General requirements for safety - Collateral standard: Programmable electrical medical systems

EN ISO14971

Medical devices - Application of risk management to medical devices

Classification

Anti-electric-shock type:

Class І with internal power supply

Anti-electric-shock degree:

Type CF with defibrillator protection

Degree of protection against harmful ingress of water:

Ordinary equipment (Sealed equipment without liquid proof)

Degree of safety of application in the presence of flammable gas:

Equipment not suitable for use in the presence of flammable gas

Working Mode:

Continuous operation

EMC:

Group І, type B

Dimensions

345mm×300mm×80mm

Net Weight

2.5kg

Display

ECG-1106: 5.7 inch monochrome LCD Screen

Environment


Transport

Storage

Working

Temperature

-20℃~+55℃

-10℃~+40℃

-5℃~+40℃

Relative Humidity

25%~95%

25%~85%

25%~85%

Atmospheric Pressure

70kPa~106kPa

70kPa~106kPa

70kPa~106kPa

Power Supply

Mains Supply

Rated input voltage =100V-240V

Rated frequency = 50/60Hz

Rated input power = 60VA

External DC power supply

Rated output voltage=12V

Rated output current=3A

Built-in Lithium Battery Pack

Rated voltage = 14.4V

Discharge cut - off voltage≥11V

Rated capacity = 2000mAh

Charge mode: Constant current/voltage

Charge current (standard) <0.2C5A(400mA)

Charge voltage (standard) = (16.8±0.1V)

Cycle life ≥ 500 times

Fuse

T2A 250V Ø5×20

Recorder

Record Mode

Thermal dot-matrix printer

Recording Paper

Rolled or folded thermo-sensitive paper

Paper Width

112mm

Effective width

104mm

Paper Speed

6.25 mm/s、12.5 mm/s25mm/s50mm/s (±3%)

HR Recognition

Technique

Peak-peak detection

HR Range

30BPM~215BPM

Accuracy

±1BPM

ECG Unit

Leads:

12 standard leads, lead change

manually/automatically

Acquisition Mode:

12 lead acquisition simultaneously

Input Circuit

Floating, protection against defibrillator effect

A/D Resolution:

12 bit

Time Constant:

≥3.2s

Frequency Response:

0.05Hz ~ 165Hz

Sensitivity:

Auto,2.5, 5, 10, 20, 40 (mm/mV) ±5 %

Input Impedance:

≥50MΩ(10Hz)

Input Circuit Current:

≤50nA

Input Voltage Range

±5 mVpp

Calibration Voltage:

1mV±2%

Depolarization Voltage

±500m V

Noise:

<15μVp-p

Filter

EMG Filter: 25/45Hz

ADS Filter: Yes/No

HUM Filter: Yes/No

CMRR

≥100dB

Patient Leakage Current:

<10μA  (a.c.)

Dielectric Strength:

4000V rms

External Input/Output

(Optional)

Input (Single ended)

Input impedance ≥100kΩ; Sensitivity10mm/V±5%;

Output(Single ended)

Output impedance≤100Ω;

Sensitivity1V/mV±5%;

Communication Interface

RS232 serial port/USB