Electocardiograph
Carewell Digital 1112M ECG
12 channel digital electrocardiograph
Quick overview
2. 7” LCD, rea-time display of 12-lead ECG waveforms
3. Detection and alarm on lead-off and low battery
4. Automatic adjustment of baseline for optimaprinting
5. Built-in ECG simulator for DEMO purpose
6. Pre-10-second printing to print out any abnormaECG waveform
7. Unique 6CH+P printing mode, convenient for physicaexamination and printing report on A4 paper
8. 3 kinds of operation mode: AUTO, MAN and ANA
9. 3 kinds of filter: HUM, EMG and ADS
10. Power supply: AC, DC and battery
11. 2 types lead mode: Standard and Cabrera
12. 1,000 ECG files can be saved
13. Soft, silicon-gekeyboard, alphameric keyboard
14. Auto-save function: printed ECG files wilbe saved in the ECG automatically
15. Communication with PC (Option)
Details
Functions of ECG-1112M are listed in below table
Model |
AUTO MEAS |
AUTO ANA |
LCD SIZE |
DISPLAY |
PAPER WIDTH |
INFO |
FILE |
HELP |
RS232 |
USB |
WIRE- LESS |
SD CARD |
ECG-1112M |
Yes |
Yes |
7” |
12CH |
210/216MM |
Yes |
1,000 |
Yes |
Yes |
Yes |
Option |
Yes |
Standard configuration
- ECG-1112M main unit
- Patient cable
- Chest electrode
- Limb electrode
- Thermal paper
- Paper roller
- Rechargeable battery
- Dust cover
Options
- PCECG-500A Workstation software package
- Aluminum trolley
Safety Standards |
MDD93/42/EEC |
Medical Device Directive |
||
IEC 60601-1: 1988 +A1:1991+A2:1995 |
Medical electrical equipment - Part 1: General requirements for safety |
|||
IEC60601-2-25: 1993 + A1:1999 |
Medical electrical equipment - Part 2-25: Particular requirements for the safety of electrocardiographs |
|||
IEC 60601-1-2: 2007 |
Medical electrical equipment Part 1: General requirements for safety -2 Collateral standard: Electromagnetic compatibility - Requirements and tests |
|||
EN 60601-1-4:1999 |
Medical electrical equipment - Part 1-4: General requirements for safety - Collateral standard: Programmable electrical medical systems |
|||
EN ISO14971:2007 |
Medical devices - Application of risk management to medical devices |
|||
Classification |
Anti-electric-shock type: |
Class І with internal power supply |
||
Anti-electric-shock degree: |
Type CF with defibrillator protection |
|||
Degree of protection against harmful ingress of water: |
Ordinary equipment (Sealed equipment without liquid proof) |
|||
Degree of safety of application in the presence of flammable gas: |
Equipment not suitable for use in the presence of flammable gas |
|||
Working Mode: |
Continuous operation |
|||
EMC: |
Group І, Class B |
|||
Dimensions |
338mm×280mm×85mm |
|||
Net Weight |
3.5kgs |
|||
Display |
ECG-1112 Class L: 5.7 inches 320×240monochrome LCD Screen |
|||
Environment |
|
Transport |
Storage |
Working |
Temperature |
-20℃~+55℃ |
-10℃~+40℃ |
+5℃~+40℃ |
|
Relative Humidity |
25%~95% |
25%~85% |
25%~85% |
|
Atmospheric Pressure |
700hPa ~1060hPa |
700hPa ~1060hPa |
700hPa ~1060hPa |
|
Power Supply |
AC power Supply |
Rated input voltage =100V~240V |
||
Rated frequency = 50/60Hz |
||||
Rated input power = 70VA |
||||
External DC power supply |
Rated output voltage=19V |
|||
Rated output current=3.5A |
||||
Built-in Lithium Battery Pack |
Rated voltage = 14.4V |
|||
Discharge cut - off voltage≥11V |
||||
Rated capacity = 4400mAh |
||||
Charge mode: Constant current/voltage |
||||
Charge current (standard) <0.3C5A(1320mA) |
||||
Charge voltage (standard) = (16.8±0.1V) |
||||
Cycle life ≥ 500 times |
||||
Power Consumption |
70VA(max) |
|||
Recorder |
Record Mode |
Thermal dot-matrix printer |
||
Recording Paper |
Rolled or folded thermo-sensitive paper |
|||
Paper Width |
210/216 mm |
|||
Effective width |
200/206mm |
|||
Paper Speed |
6.25 mm/s、12.5 mm/s、25mm/s、50mm/s (±3%) |
|||
HR Recognition |
Technique |
Peak-peak detection |
||
HR Range |
30BMP~215BMP |
|||
Accuracy |
±1 BPM |
|||
ECG Unit |
Leads: |
12 standard leads, lead change manually/automatically |
||
Acquisition Mode: |
12 lead acquisition simultaneously |
|||
Input Circuit |
Floating, protection against defibrillator effect |
|||
A/D Resolution: |
12 bit |
|||
Time Constant: |
≥3.2s |
|||
Frequency Response: |
0.05Hz ~ 165Hz |
|||
Sensitivity: |
Auto,2.5, 5, 10, 20,40 (mm/mV) ±5 % |
|||
Input Impedance: |
>50MΩ(10Hz) |
|||
Input Circuit Current: |
≤50nA |
|||
Input Voltage Range |
±5 mVpp |
|||
Calibration Voltage: |
1mV±2% |
|||
Depolarization Voltage |
±500m V |
|||
Noise: |
<15μVp-p |
|||
Filter |
EMG Filter: 25/45Hz |
|||
ADS Filter: Yes/No |
||||
HUM Filter: Yes/No |
||||
CMRR |
≥100dB |
|||
Patient Leakage Current: |
<10μA (a.c.) |
|||
Dielectric Strength: |
4000V rms |
|||
External Input/Output (Optional) |
Input (Single ended) |
Input impedance≥100kΩ; Sensitivity10mm/V±5%; |
||
Output (Single ended) |
Output impedance≤100Ω; Sensitivity1V/mV±5%; |
|||
Communication Interface |
RS232 serial port/USB |
More products
-
Carewell Digital-1103L/1103LW ECG
ECG-1103L/1103LW
Three channel digital electrocardiograph -
Carewell Digital-1106L ECG
ECG-1106L
Six channel digital electrocardiograph -
Carewell Digital T12 ECG
T12
12 channel digital electrocardiograph